Mechanical ventilator control unit design with a Venturi-type flowmeter

,

1Halic University,15 Temmuz Şehitler Caddesi, No:14/12 34060 Eyüpsultan/Istanbul

2Beykent University, Ayazağa, Hadım Koruyolu Cd. No:19, 34398 Sarıyer/İstanbul

Published
16 February 2023
Pages
340-348
Received
16 February 2023
DOI
10.65582/set.2022.084
Access
Open access, © 2022 Copyright WSSET
Download PDF9 pages · 784 KB

Abstract

The importance of the mechanical ventilators has increased as their presence became more of an issue during the COVID-19 pandemic. The air/oxygen control unit is the vital element of the machine that adjusts the air quality according to the patient’s needs. For this purpose, design optimization of a mechanical ventilator control unit was performed. The design procedure was conducted by taking into account safety and flow adjustment considerations. Furthermore, a Venturi-type flowmeter was designed and embedded into the main body of the ventilator. Flow measurements using a Venturi flowmeter were achieved successfully. Afterwards, a computational fluid dynamics study was carried out to estimate the flow behaviour at varying valve opening conditions. Also, a numerical noise prediction study was accomplished based on Broadband Noise Source Model. Finally, the effect of inlet pressure on flow rate was investigated with a parametric study, and required inlet pressure was determined. The results showed that the Venturi meter with a 4mm throat diameter at 0.5 bar inlet pressure condition satisfied the desired operating conditions.

Keywords. mechanical ventilator; air/oxygen control unit; COVID-19; noise prediction; computational fluid dynamics

© 2022 Copyright WSSET

Free to read, download, print and share, unaltered and for non-commercial purposes, provided the authors, the paper and the proceedings are credited. Any other use, including commercial use or adaptation, needs written permission: write to [email protected].

← SET Proceedings